AED7000 plus is a portable model, which can be equipped at home, in public places, or in the hospitals. It’s easy and convenient to use when give first-aid to patient. Meanwhile, it has the function of automatic analyzing patient’s ECG data, and then takes corresponding defibrillation energy level according to patient’s current situation, which has greatly improved success rate, and maximum reduced damage to patient’s heart.
Automatic External Defibrillator Features:
1. Three-step defibrillation process
2. Two-button operation
3. Extensive voice and visual prompts for the operator
4. Biphasic energy output
5. Lock-out protection to prevent inadvertent defibrillation
6. Continuous event recording for reporting each use to a printer or computer
7. Weekly self-test to ensure readiness
8. Color LCD screen
9. Adult and Pediatric Defibrillation
10. CPR coaching(Optional)Automatic External Defibrillator Specification:
1. Three-step defibrillation process
2. Two-button operation
3. Extensive voice and visual prompts for the operator
4. Biphasic energy output
5. Lock-out protection to prevent inadvertent defibrillation
6. Continuous event recording for reporting each use to a printer or computer
7. Weekly self-test to ensure readiness
8. Color LCD screen
9. Adult and Pediatric Defibrillation
10. CPR coaching(Optional)Automatic External Defibrillator Specification:
Category
|
Specification
|
Output
|
Biphasic Truncated exponential
|
Energy Sequence(adult )
|
150/150/200(Default)
|
150/200/200
|
|
200/200/200
|
|
Note: the Energy sequence can be changed using the computer IR port.
|
|
Energy Sequence(Pediatric)
|
50/70/85(Default)
|
70/70/85
|
|
70/85/85
|
|
85/85/85
|
|
Note: the Energy sequence can be changed using the computer IR port.
|
|
Charge Time
|
10 sec. to 200 J
Lithium ion manganese oxide battery (LMO)
|
Analysis Time
|
9 sec.
|
The maximum time from the initiation of rhythm analysis to readiness for maximum Energy discharge after 15 shocks.
|
Less than 30 seconds
|
The maximum time from initially switching power on to readiness for maximum Energy discharge after 15 shocks..
|
Less than 35seconds
|
Audible Prompts
|
19 audible prompts
|
Visual Prompts
|
LED/LCD
|
Controls
|
six buttons – On/Off, Shock, and 4 software-configurable buttons
|
Output Energy Accuracy
|
±15% into any impedance from 25 to 175Ω
|
The Maximum Voltage
|
1800±20V
|
Output disabled when the PATIENT impedance is outside limits
|
20Ω to 200Ω
|
Recovery of the ECG input after defibrillation
|
<10s
|
the time from switching “ON” until the AED7000 Plus is ready for NORMAL USE,
|
<4s
|
he time required for the ME EQUIPMENT to warm from the minimum storage temperature between uses (4.2.1) until the ME EQUIPMENT is ready for its INTENDED USE when the ambient temperature is 20 °C
|
>30min
|
he time required for the ME EQUIPMENT to cool from the maximum storage temperature between uses (4.2.1) until the ME EQUIPMENT is ready for its INTENDED USE when the ambient temperature is 20 °C
|
>1min
|