Syringe Pump
Model: BD-5000 technical specification:
Maxim. Injection rate: 1500 ml/h (Different syringes have different maximum rate)
Applicable syringes: All brands of standard syringes: 5, 10, 20, 30 and 50 (60) ml
Drug library: More than 20 drug names: Penicillin, Vancomycin, Pitocin, Insulin, Lidocaine, Heparin, Atropin, Dopamine, Epinephrine etc.
Power supply: AC100-240V.50/60HZ
Gross weight: <3.0kg
Battery: Rechargeable lithium polymer battery 7.4V, 1600mA working for 4 hours (5ml/h) after being fully charged
Max power consumption: 25VA
Accumulated volume: 0.1~9999.9ml
Injection accuracy: ±3% (after being well adjusted, up to ±2%)
Injection increment: 0.1ml/h
Preset range: 0.1~9999.9ml
Classification: Class II, type BF
Injection rate range: 5ml: 0.1ml/h—100ml/h 10ml: 0.1ml/h—200ml/h
20ml: 0.1ml/h—400ml/h 30ml: 0.1ml/h—600ml/h
50 (60) ml: 0.1ml/h—1500ml/h
Bolus rate: 5ml: 100ml/h 10ml: 200ml/h 20ml: 400ml/h
30ml: 400-600ml/h 50ml: 400-1500ml/h
Alarm function: NEAR, OVER, OCCL., syringe disengaged, no AC power supply, low battery, serious low battery, error 1/2/3, preset volume alarm
Selectable modes: 1. Rate Mode 2. Time Mode 3. Body Weight Mode
Time mode: Volume limit: 0.1-999.9ml Time: 1-2000min
Body weight mode: Body weight: 0.1~300kg Drug limit: 0.1~999.9mg
Volume limit: 0.1~999.9ml Dose: 0.1~9999.9(subject to unit, dosage, liquid, weight and syringe)Unit: mg/kg/h, ug/kg/min
Occlusion value: 0.04Mpa-0.16Mpa
KVO rate: Adjustable range: 0.1ml/h-5.0ml/h, choose “OFF” to cancel
EMC: Meet the requirements of IEC60601-1-2(2007 version)
Electrical safety: Meet the requirements of EN60601-1:1990+A1:1993+A2:1995
Two CPU