KEY FEATURES
40.68 MHz Radio frequency sealing – The radio frequency waves generated by an optimally designed RF oscillator guarantee exceptional seal quality.
No risk of haemolysis of blood in the tube segments as the radio frequency sealing heats only the tube and not the blood inside.
The RF signal is applied only after the tube is fully squeezed.
Easily accessible electrodes for cleaning and disinfecting.
A sealer comes equipped with SMPS to withstand wide input voltage fluctuations, which ensures that the performance of equipment remains constant irrespective of the input voltage, ensuring a consistent seal quality.
SPECIFICATIONS
Power supply input frequency
100 V AC to 240 V AC, 50 Hz/60 Hz
Safety standard
IEC 61010-1;2010
Electrical safety
Protection against electric shock Class I
Operation mode
Once it is turned on, the device is ready to use
RF output power (effective)
20 W (max)
Indicator lights
Cover open (yellow), Power (green), Ready (green), Seal (red)
Tube detection
Automatic
RF output frequency
40.68 MHz
Tube specification
Polyvinyl chloride (PVC) tubing up to 6 mm (outer daimeter)
Sealing time
1 to 3 seconds, adjustable
Dimensions* (w x d x h)
155 mm x 330 mm x 160 mm
Weight*
5.25 kg
Power consumption
Operating: 185 W (max)
Standby: 20 W (max)
Fuse
250 V, 3.5 A, type T, slow blow
International compliance
Low Voltage Directive (CE mark), China
RoHS, European Restriction of Hazardous Substances
Manufacturing standard
ISO 9001:2008, EN ISO 13485:2012