Infusion Pump DI-2000

MEDIFUSION Volumetric Infusion Pump

FEATURES

  • Compact Design
  • Small size – 1.7kg
  • Compatible with various kinds of IV set
  • Brdad range of flowrate : 0.1 to 1,299 ml/h
  • One touch Bolus, Pause function
  • Dosage-auto calculation (Input data : boby weight, drug volume, Solution volume, dose rate)
  • Titration – Enable to change the flowrate during infusion
  • Keylock for safety
  • Event history up to 2000
  • Firmware upgrade via internet (system or software upgrade)
  • Quality Assurance & Certification – ISO13485, CE
Pumping Mechanism Peristaltic Transit Finger
Flow Rate 0.1 ~ 1200 ml/h Micro (On): 0.1 ~ 999.9 ml/h (in 0.1 ml/h increments) 1000 ~ 1200 ml/h (in 1 ml/h increments) Micro (Off): 1 ~ 1200 ml/h (in 1 ml/h increments)
Accuracy ±5% (BD I.V Set, AN122) Measured every hour at room temperature (25℃) using water or saline solution (above 1.0 ml/h of flow rate).

* Accuracy may differ depending on the type of medical fluid and IV Set & Feeding Set.

Infusion Range Total Vol. 0.1 ~ 9999 ml or infinity Micro (On): 0.1 ~ 999.9 ml (in 0.1 ml increments) 1000 ~ 1200 ml (in 1 ml increments) Micro (Off): 1 ~ 9999 ml (in 1 ml increments)
Infused Vol. 0.0 ~ 9999 ml 0.0 ~ 999.9 ml (in 0.1 ml increments) 1000 ~ 9999 ml (in 1 ml increments)
Bolus Rate (Purge) 700 ml/h (default) – 1 ~ 1200 ml/h
Bolus Volume (Purge) 5 ml (default) – 1 ~ 9999 ml
KVO Rate 1 ml/h (default) 1 ~ 9 ml/h At flow rate 1.0 ~ 1200 ml/h
0.1 ml/h (default) At flow rate 0.1 ~ 0.9 ml/h
Occlusion pressure 100mmHg ~ 950mmHg(13~126kPa)
Display 7 segments (4 digits * 3 lines)
Alarms * Air Bubble alarm

* Occlusion alarm (detection capacity: 100 ~ 950 mmHg) – 9 adjustable steps for downstream occlusion

* Door Open alarm

* Battery Low, Battery Empty alarm

* Abnormal Drop alarm

* Infusion Complete alarm (automatic switch-over to KVO feature)

* AC/DC Power Shut-off alarm

* Start Reminder alarm (pause set time elapsed)

* Standby alarm (at an interval of 2 minutes when not infusing)

* Near complete alarm

* The pump stops operating when alarms go off except for battery low, start reminder and infusion completion signals.

* Malfunction – The status of the pump is displayed and an alarm goes off.

Safety Features * Door Open: Infusion is unavailable when the door is open.

* Key Lock: Only the START, STOP and ON/OFF keys work.

* Air Sensor: Detects air bubbles.

* Occlusion Sensor: Detects clogging of the tube.

* Drop Sensor: Detects medical fluid drops.

* Nurse Call: Monitoring is possible if an alarm goes off.

Other Functions * Time Mode: Calculates the flow rate by setting the infusion time.

* Gtt Mode: Calculates the flow rate by setting the number of infusion drops.

* Dosage Mode: Calculates the flow rate by setting the dosage.

* Titration: Flow rate can be altered during infusion (available only when the remaining time for infusion is more than 5 minutes).

* Battery Low, Empty: Alarms go off 30 minutes and 3 minutes before the battery goes flat.

* Date and Time: The date and time can be checked without a power supply, since a clock is installed.

* Pause: Infusion resumes automatically after preset time elapses when the pause key is pressed and infusion is stopped (default 24 hours, setting range from 1 minute to 24 hours in 1 minute units).

* Event History: Saves up to 2000 and it can view by separate PC utility. Please refer service manual for detailed usage of event history.

* Alarm log event: Saves up to 50 in the pump and it can view in the pump at setting mode1. Please refer service manual for detailed usage of alarm log event.

* The electronic data of the infusion pump are stored for over 2 years even if it has been turned off.

Power * AC 100~240 VAC, 50/60 Hz (Fuse: 250 V, T3.15 A)

* DC 12V (500 mA)

* Ni-MH Rechargeable Battery – 2,000 mA/h, AA, 1.2V x 10cell – operating time – approx. 6 hours at a flow rate of 25 ml/h – recharging time – approx. 5 hours

Power Requirements 35VA
Protection Type * Protection from electric shock: Class â… 

* Protection capacity for electric shock: Type CF Protection from liquid infiltration: IPX 2 (drip-proof)

Size (D×W×H) 120×130×206 mm
Weight approx. 1.7 kg
Operating Environment Temperature: 10~ 40℃, Humidity: 20 ~ 90% RH Atmospheric Pressure: 70 ~ 106 kPa (525.04 ~ 795.06 mmHg)
Storage Environment Temperature: -10 ~ 60℃, Humidity: 10 ~ 95% Atmospheric Pressure: 65 ~ 120 kPa (487.54 ~ 900.07 mmHg)