BASE UNIT
DIMENSIONS (W X H X D)
Standard Trolley Version (Cart) with COSY with 2 Vaporizers Approx. 35.8 x 56.7 x 30.3 in (91 x 144 x 77 cm)
High Trolley Version (Cart) with COSY with 2 Vaporizers Approx. 35.8 x 59 x 30.3 in (91 x 150 x 77 cm)
Wall / Ceiling Version with COSY with 2-Vaporizers Approx. 33.5 x 30.7 x 19.7 in (85 x 78 x 50 cm)
WEIGHT AND LOAD
Standard Trolley (with COSY)
without supplementary cylinders and vaporizers
297,6 lb (135 kg)
High Trolley (with COSY)
without reserve-gas cylinders and vaporizers
308,7 lb (140 kg)
Wall and Ceiling Mount (with COSY and without supplementary
cylinders and vaporizers)
152.1 lb (69 kg)
POWER AND BATTERY BACKUP
Power Input 100 to 240 VAC, 50/60 Hz, 70 VA
Operation time with fully charged batteries 45 minutes minimum (up to 120 minutes)
ANAESTHESIA GAS SUPPLY MODULE
Range of fresh gas flow indicators 0.0 to 12.0 L/min
O2 flush max. 75 L/min at 87 psi (6 kPa x 100)
Vapor mount Up to two Dräger (Auto Exclusion/Interlock) or Selectatec® Vapors
optional additional Dräger (Auto Exclusion/Interlock) or Selectatec
Vapor standby holder
VENTILATOR OPERATING SPECIFICATIONS
Ventilator E-vent® Electronically controlled, electrically driven
Operating modes Volume Controlled Ventilation
Pressure Controlled Ventilation (Optional)
Pressure Support (Optional)
SIMV/PS (Optional)
Manual Ventilation
Spontaneous Breathing
CONTROL INPUT RANGES
Breathing Frequency (rate) 4 to 60 bpm (1/min)
Positive End Expiratory Pressure (PEEP) 0 to 20 cmH2O (hPa)
Inspiration/expiration ratio (Ti:Te) 4:1 to 1:4
Pressure limiting (Pmax) 15 to 70 cmH2O (hPa)
Tidal Volume (Vt) 20 to 1,400 mL in Volume Control
20 to 1,100 mL in SIMV/PS
Inspiration pause (Tip:Ti) 0 to 50 %
SIMV Inspiratory time 0.3 – 4.0 sec
Inspiratory pressure (Pinsp) PEEP + 5 to 65 cm H2O (hPa)
Inspiratory Flow (InspFlow) 10 to 75 L/min in Volume and Pressure Control modes
10 to 85 L/min in Pressure Support and SIMV/PS modes
Pressure Support Level (PPS) PEEP +3 to 20 cmH2O (hPa)
Min. frequency for apnoe-ventilation (Freq. Min.) 3 to 20 bpm and “OFF”
Trigger level 2 to 15 L/min
Technical Data
Integrated Safety Functions
Sensitive Oxygen Ratio Controller (S-ORC) guarantees a minimum
O2 concentration of 23% in an O2 /N2O mixture.
N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than
0.2 L/min. Audible and visual (flashing red LED) indication in case
O2 pressure drops below 1.38 bar (20 psi) ± 0.27 bar (4 psi).
In case of electricity and battery failure, manual ventilation, gas
delivery and agent delivery are possible. High Pressure Safety
Relief Valve opens at 75 ± 5 cmH2O (hPa).
Negative Pressure Safety Relief Valve opens at -8 ± 2 cmH2O (hPa).
VENTILATOR MONITORING
Monitoring
Continuous monitoring of inspiratory O2 concentration (optional),
breathing frequency, tidal volume expiration, expiratory minute
volume, peak airway pressure, PEEP, and selection of mean or
plateau pressure.
Expiratory Minute Volume Range
0 to 99 L/min
Control Screen
10.4 in (26,4 cm) TFT color screen, day and night mode
BREATHING SYSTEM
Optional breathing system heater yes
Volume of entire compact breathing system, with absorber, excl. hoses, excl. bag 1.7 L
Volume of CO2 absorber Standard: 1.5 L reusable canister, Option: 1.2 L prefilled Drägersorb CLIC absorber
Mounting on swivel arm, left or right, height adjustable
GAS SUPPLY AND CONNECTION
Gas Supply Versions O2/ N2O/ Air, O2/, Air, O2/ N2O
Cylinder Supply Versions (optional) O2, O2/ Air, O2/ N2O, 1 or 2 standing gas cylinders, pin-index
hanger yokes for 2 cylinders
OTHER
Communication Interface 1 RS-232 standard, additional RS-232 optional
Protocols Vitalink and MEDIBUS
Data available for Export All alarms, pressure, O2 and volume data, ventilation settings, flow
curve and pressure curve
Power outlets 2 integrated, medical grade power outlets with circuit breakers
(option)
Writing surfaces Pull-out tray (standard), Flip-up tray (optional)
Additional accessories Anaesthetic gas scavenging system (AGSS), endotracheal suction
unit, integrated auxiliary oxygen flowmeter, auxiliary common gas
outlet (CGO), low flow kit, power outlet strip
Mounting options Mounting rails on side and rear panels, side/ top mountings for
patient monitor, patient gas measurement module, infusion pumps,
infusion bags, storage basket